The use of collimator angle optimization and jaw tracking for VMAT‐based single‐isocenter multiple‐target stereotactic radiosurgery for up to six targets in the Varian Eclipse treatment planning system
Journal of Applied Clinical Medical Physics2021Vol. 22(9), pp. 171–182
Citations Over TimeTop 14% of 2021 papers
Lauren M. M. Pudsey, Dean Cutajar, A. Wallace, Anastasia Saba, Laurel Schmidt, Andrej Bece, C. Clark, Yoshiya Yamada, Giordano Biasi, Anatoly Rosenfeld, Joel Poder
Abstract
These findings suggest that CAO has limited benefit in VMAT SIMT SRS of 2-6 targets when using a low-dose penalty to the healthy brain during plan optimization in Eclipse. As clinical benefits of jaw tracking were found to be minimal and plan deliverability was potentially reduced, a cautious approach would be to exclude jaw tracking in SIMT SRS plans.