Advanced c‐myc antisense (AVI‐4126)‐eluting phosphorylcholine‐coated stent implantation is associated with complete vascular healing and reduced neointimal formation in the porcine coronary restenosis model
Citations Over TimeTop 10% of 2004 papers
Abstract
An advanced six-ring morpholino backbone c-myc antisense (AVI-4126) was shown to inhibit c-myc expression and intimal hyperplasia after local catheter delivery in a porcine balloon injury model. The purpose of this study was to investigate the effects of an AVI-4126-eluting phosphorylcholine-coated (PC) stent on c-myc expression restenosis and vascular healing after stent implantation in porcine coronary arteries. PC stents were loaded with AVI-4126 using soak trap. Nine pigs underwent AVI-4126 PC coronary stent implantation (two stents/animal). Two to six hours postprocedure, three pigs were sacrificed and stented segments were analyzed by Western blot for c-myc expression. In chronic experiments, six pigs (12 stent sites) were sacrificed at 28 days following intervention and vessels were perfusion-fixed. High-performance liquid chromatography analysis of plasma samples showed minimal presence of the antisense. Western blot analysis of the stented vessels demonstrated inhibition of c-myc expression at 2 and 6 hr after procedure. Quantitative histologic morphometry showed that the neointimal area was significantly reduced (by 40%) in the antisense-coated group compared with control (2.3 +/- 0.7 vs. 3.9 +/- 0.8 mm(2), respectively; P = 0.0077). Immunostaining and electron microscopy demonstrated complete endothelialization, without fibrin deposition, thrombosis, or necrosis in all implanted stents. In the porcine coronary model, an advanced c-myc-eluting PC stent blocked c-myc expression and significantly inhibited myointimal hyperplasia and allowed complete reendothelialization and healing response.
Related Papers
- → Optical coherence tomography findings of very late stent thrombosis after drug-eluting stent implantation(2011)27 cited
- → miR‐29a regulates vascular neointimal hyperplasia by targeting YY1(2016)13 cited
- → Optical Coherence Tomographic Observation of Morphological Features of Neointimal Tissue after Drug-Eluting Stent Implantation(2014)7 cited
- → RING finger protein 10 prevents neointimal hyperplasia by promoting apoptosis in vitro and in vivo(2018)5 cited
- → Effective Intraluminal Delivery of TSP-2 SiRNA to Reduce Intimal Hyperplasia in a Rat Model(2013)