Portable perfusion phantom for quantitative DCE‐MRI of the abdomen
Medical Physics2017Vol. 44(10), pp. 5198–5209
Citations Over TimeTop 12% of 2017 papers
Harrison Kim, Mina Mousa, Patrick Schexnailder, Robert W. Hergenrother, Mark Bolding, Bernard Ntsikoussalabongui, Vinoy Thomas, Desiree E. Morgan
Abstract
The perfusion phantom reduced variability in quantitating contrast concentration and Ktrans values of human abdominal tissues across different MR units, but static phantom did not. The perfusion phantom has the potential to facilitate multi-institutional clinical trials employing quantitative DCE-MRI to evaluate various abdominal malignancies.