Experimental investigation of GafChromic® EBT3 intrinsic energy dependence with kilovoltage x rays, 137Cs, and 60Co
Medical Physics2017Vol. 45(1), pp. 448–459
Citations Over TimeTop 10% of 2017 papers
Abstract
For the beam qualities investigated, a single energy-dependent kbq correction can be used for doses between 500 and 7002 mGy. Using the single-hit model with the single-free-parameter fit to solve for β shows promise in the determination of the intrinsic energy response of film, with β being the mathematical analog of the measured kbq .
Related Papers
- → Practical calibration curve of small-type optically stimulated luminescence (OSL) dosimeter for evaluation of entrance skin dose in the diagnostic X-ray region(2015)39 cited
- → A pilot study of a postal dosimetry system using the Fricke dosimeter for research irradiators(2021)9 cited
- → Absorbed dose to water based dosimetry versus air kerma based dosimetry for high-energy photon beams: an experimental study(2002)20 cited
- → A CaS:Ce,Sm-based dosimeter for online dosimetry measurement(2011)3 cited
- → The impact of field size and radiation quality on KAP-meter and CT-chamber response(2018)1 cited