Evaluation of machine log files/MC‐based treatment planning and delivery QA as compared to ArcCHECK QA
Medical Physics2018Vol. 45(7), pp. 2864–2874
Citations Over TimeTop 16% of 2018 papers
Carl Stanhope, Douglas G. Drake, Jian Liang, M. Alber, Matthias Söhn, Charbel Habib, Virgil Willcut, Di Yan
Abstract
Reconstructing every log file sample with no increase in computational cost, MC-based LF QA is faster and more accurate than CS-based LF QA. Offering similar dosimetric accuracy compared to AC measurement, MC-based log files can be used for treatment planning QA.