Increase of the transmission and emittance acceptance through a cyclotron‐based proton therapy gantry
Medical Physics2022Vol. 49(4), pp. 2183–2192
Citations Over TimeTop 10% of 2022 papers
Vivek Maradia, Anna Chiara Giovannelli, D. Meer, Damien C. Weber, Antony Lomax, Jacobus Maarten Schippers, S. Psoroulas
Abstract
We have developed a new gantry beam optics which, by selecting a large beam size and low divergence at the gantry entrance and using an imaging factor of 0.5 (2:1), increases the emittance acceptance of the gantry, leading to a substantial increase in beam intensity at low energies. We expect that this approach could easily be adapted for most types of existing gantries.