A nonconvex model‐based combined geometric calibration scheme for micro cone‐beam CT with irregular trajectories
Medical Physics2023Vol. 50(5), pp. 2759–2774
Citations Over TimeTop 11% of 2023 papers
Abstract
The proposed combined geometrical calibration scheme does not rely on a phantom with an intricate pattern of fiducials, so it is applicable in Micro CT with high resolution. The two parts of the scheme, the "measurement model" and the "calibration model," prove to be of high accuracy. The combination of these two models can effectively improve the calibration accuracy, especially in some extreme cases.