Patient‐specific quality assurance of dynamically‐collimated proton therapy treatment plans
Medical Physics2024Vol. 51(9), pp. 5901–5910
Citations Over TimeTop 22% of 2024 papers
Laura C. Bennett, Daniel E. Hyer, Justin Vu, Kaustubh Patwardhan, Kevin Erhart, Alonso N. Gutiérrez, Eduardo Pons, Eric Jensen, Manual Ubau, Julio Zapata, Andrew Wroe, Karsten K. Wake, Nicholas P. Nelson, Wesley S. Culberson, Blake R. Smith, Patrick M. Hill, Ryan T. Flynn
Abstract
The PSQA process has been validated and experimentally verified for DCS-collimated PBS. Dosimetric agreement between the measured and calculated doses was demonstrated to be similar for DCS-collimated PBS to that achievable with noncollimated PBS.
Related Papers
- → Contour scanning, multi-leaf collimation and the combination thereof for proton pencil beam scanning(2018)18 cited
- → Dosimetric delivery validation of dynamically collimated pencil beam scanning proton therapy(2023)8 cited
- → Integration and dosimetric validation of a dynamic collimation system for pencil beam scanning proton therapy(2023)3 cited
- → TH-C-BRD-03: Determining the Optimal Collimator Position for Collimated Pencil Beam Scanning Proton Therapy(2014)
- Beam collimation for CSNS/RCS(2010)