Signal‐to‐noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coil
Citations Over TimeTop 10% of 2001 papers
Abstract
Calculations of the RF magnetic (B(1)) field as a function of frequency between 64 and 345 MHz were performed for a head model in an idealized birdcage coil. Absorbed power (P(abs)) and SNR were calculated at each frequency with three different methods of defining excitation pulse amplitude: maintaining 90 degrees flip angle at the coil center (center alpha = pi/2), maximizing FID amplitude (Max. A(FID)), and maximizing total signal amplitude in a reconstructed image (Max. A(image)). For center alpha = pi/2 and Max. A(image), SNR increases linearly with increasing field strength until 260 MHz, where it begins to increase at a greater rate. For these two methods, P(abs) increases continually, but at a lower rate at higher field strengths. Above 215 MHz in MRI of the human head, the use of FID amplitude to set B(1) excitation pulses may result in apparent decreases in SNR and power requirements with increasing static field strength. Magn Reson Med 45:684-691, 2001.
Related Papers
- → Specific absorption rate and temperature distributions in human head subjected to mobile phone radiation at different frequencies(2011)103 cited
- → Specific Absorption Rate and Temperature Distributions in the Human Head with Implanted Deep Brain Stimulation Subjected to Mobile Phone Electromagnetic Radiation(2023)8 cited
- → Electromagnetic radiation reduction using novel metamaterial for cellular applications(2020)24 cited
- → SAR assessment in a human head model exposed to radiation from mobile phone using FEM(2003)8 cited
- → Reduction of specific absorption rate in human tissues using split ring resonators(2009)4 cited