Improved spatial harmonic selection for SMASH image reconstructions
Citations Over TimeTop 24% of 2001 papers
Abstract
The fitting of coil sensitivity functions to spatial harmonics is central to image reconstructions using the simultaneous acquisition of spatial harmonics (SMASH) technique. It has previously been shown that the selection of the set of spatial harmonics used in a SMASH reconstruction can have a noticeable effect on the quality of the reconstructed image. However, a mechanism for automatic selection of the best set of harmonics in any particular situation has not been provided. In this work, a modification to the SMASH reconstruction procedure is introduced that allows the use of a weighted average of all possible harmonics in a reconstruction. The new reconstruction procedure is shown to allow automatic selection of the spatial harmonics and substantially improve SNR for both phantom and in vivo images.