Fast‐kzthree‐dimensional tailored radiofrequency pulse for reducedB1inhomogeneity
Magnetic Resonance in Medicine2006Vol. 55(4), pp. 719–724
Citations Over TimeTop 10% of 2006 papers
Abstract
This article presents a small-flip-angle, three-dimensional tailored RF pulse that excites thin slices with an adjustable quadratic in-plane spatial variation. The quadratic spatial variation helps to compensate for the loss in image uniformity using a volume coil at 3 T due to the wavelike properties of the RF field. The pulse is based on a novel "fast-kz" design that uses a series of slice-select subpulses along kz and phase encoding "blips" along kx-ky. The method is demonstrated by acquiring a series of 5-mm-thick T2-weighted images of the human brain at 3 T using pulses 4.8 ms in length with a 45 degrees flip angle.
Related Papers
- → Characterization and evaluation of a flexible MRI receive coil array for radiation therapy MR treatment planning using highly decoupled RF circuits(2018)45 cited
- → A Preliminary Study for Reference RF Coil at 11.7 T MRI: Based on Electromagnetic Field Simulation of Hybrid-BC RF Coil According to Diameter and Length at 3.0, 7.0 and 11.7 T(2022)10 cited
- → Analytical approach to noncircular section birdcage coil design: Verification with a Cassinian oval coil(2004)15 cited
- Radiofrequency coil in rat brain MR imaging(2010)
- Fabrication of a Staircase Coil with Improved SNR and Image Uniformity by Structural Changes of a Conventional Birdcage Coil at 1.5T MRI(2003)