On the accuracy of T1 mapping: Searching for common ground
Magnetic Resonance in Medicine2014Vol. 73(2), pp. 514–522
Citations Over TimeTop 1% of 2014 papers
Abstract
We conclude that the three most common T1 mapping protocols produce stable T1 values in phantoms, but not in vivo. To improve the accuracy of T1 mapping, we recommend that sites perform in vivo validation of their T1 mapping method against the inversion recovery reference method, as the first step toward developing a robust calibration scheme.