Magnetic resonance conditionality of abandoned leads from active implantable medical devices at 1.5 T
Magnetic Resonance in Medicine2021Vol. 87(1), pp. 394–408
Citations Over TimeTop 10% of 2021 papers
Yu Wang, Ran Guo, Wei Hu, Jianfeng Zheng, Qingyan Wang, Jay Jiang, Krishna K. N. Kurpad, Norbert Kaula, Stuart A. Long, Ji Chen, Wolfgang Kainz
Abstract
For the three investigated AIMDs under 1.5T MR scanning, our results indicate that RF-induced lead-tip heating of abandoned leads strongly depends on the proximal lead termination. A metallic cap applied to the proximal termination of the tested leads could significantly reduce the RF-induced lead-tip heating.