Performance of receive head arrays versus ultimate intrinsic SNR at 7 T and 10.5 T
Magnetic Resonance in Medicine2024Vol. 92(3), pp. 1219–1231
Citations Over TimeTop 10% of 2024 papers
Bei Zhang, Jerahmie Radder, Ilias I. Giannakopoulos, Andrea Grant, Russell Lagore, Matt Waks, Nader Tavaf, Pierre‐François Van de Moortele, Gregor Adriany, Alireza Sadeghi‐Tarakameh, Yiğitcan Eryaman, Riccardo Lattanzi, Kâmil Uǧurbil
Abstract
Major uiSNR gains are predicted for human head imaging in going from 7 T to 10.5 T, ranging from ˜2-fold at locations corresponding to the edge of the brain to 2.6-fold at the center, corresponding to approximately quadratic increase with the magnetic field. Realistic 31- and 63-channel receive arrays, however, approach the central uiSNR at 7 T, but fail to do so at 10.5 T, suggesting that more coils and/or different type of coils will be needed at 10.5 T and higher magnetic fields.