Orientation‐independent quantification of macromolecular proton fraction in tissues with suppression of residual dipolar coupling
Citations Over TimeTop 15% of 2024 papers
Abstract
Quantitative magnetization transfer (MT) imaging enables noninvasive characterization of the macromolecular environment of tissues. However, recent work has highlighted that the quantification of MT parameters using saturation radiofrequency (RF) pulses exhibits orientation dependence in ordered tissue structures, potentially confounding its clinical applications. Notably, in tissues with ordered structures, such as articular cartilage and myelin, the residual dipolar coupling (RDC) effect can arise owing to incomplete averaging of dipolar-dipolar interactions of water protons. In this study, we demonstrated the confounding effect of RDC on quantitative MT imaging in ordered tissues can be suppressed by using an emerging technique known as macromolecular proton fraction mapping based on spin-lock (MPF-SL). The off-resonance spin-lock RF pulse in MPF-SL could be designed to generate a strong effective spin-lock field to suppress RDC without violating the specific absorption rate and hardware limitations in clinical scans. Furthermore, suppressing the water pool contribution in MPF-SL enabled the application of a strong effective spin-lock field without confounding effects from direct water saturation. Our findings were experimentally validated using human knee specimens and healthy human cartilage. The results demonstrated that MPF-SL exhibits lower sensitivity to tissue orientation compared with R 2 $$ {R}_2 $$ , R 1 ρ $$ {R}_{1\rho } $$ , and saturation-pulse-based MT imaging. Consequently, MPF-SL could serve as a valuable orientation-independent technique for the quantification of MPF.
Related Papers
- → Improved MR angiography: magnetization transfer suppression with variable flip angle excitation and increased resolution.(1994)131 cited
- → Modeling the influence of TR and excitation flip angle on the magnetization transfer ratio (MTR) in human brain obtained from 3D spoiled gradient echo MRI(2010)74 cited
- → New MRI method with contrast based on the macromolecular characteristics of tissues(2003)21 cited
- → Optimization of on‐resonant magnetization transfer contrast in coronary vein MRI(2010)6 cited
- Magnetization transfer imaging of the head and neck: normative data.(1994)