Creation and comparison of MS/MS spectral libraries using quadrupole ion trap and triple‐quadrupole mass spectrometers
Citations Over TimeTop 10% of 2004 papers
Abstract
Searchable libraries of MS/MS spectra, obtained using liquid chromatography/tandem mass spectrometry (LC/MS/MS) with data-dependent scan mode switching on both quadrupole ion trap and triple-quadrupole mass spectrometers in conjunction with electrospray ionization, are presented. The effects on library search scores of changing the parameters for producing collision-induced dissociation (CID) on both instrument types are systematically evaluated. These observations serve as a basis for determining a universal set of conditions for building MS/MS libraries. A group of 19 closely related steroids was used. The ability to obtain library-searchable spectra at low concentrations is demonstrated for the analysis of a sample of progesterone spiked with hydroxyprogesterone impurities at 0.1 and 0.01%.
Related Papers
- → Triple quadrupole linear ion trap mass spectrometer for the analysis of small molecules and macromolecules(2004)284 cited
- → High-performance liquid chromatography–tandem mass spectrometry with a new quadrupole/linear ion trap instrument(2003)87 cited
- → Application of ion trap technology to liquid chromatography/mass spectrometry quantitation of large peptides(2008)43 cited
- → Multistage Fragmentation of Ion Trap Mass Spectrometry System and Pseudo-MS3of Triple Quadrupole Mass Spectrometry Characterize Certain (E)-3-(Dimethylamino)-1-arylprop-2-en-1-ones: A Comparative Study(2014)4 cited
- → Triple Quadrupole and Quadrupole Ion Trap Mass Spectrometers(2012)1 cited