Post‐interface signal suppression, a phenomenon observed in a single‐stage Orbitrap mass spectrometer coupled to an electrospray interfaced liquid chromatograph
Citations Over TimeTop 10% of 2010 papers
Abstract
Signal suppression is a common issue when analyzing compounds by liquid chromatography coupled to mass spectrometry (LC/MS/MS). Suppression of signals is caused by co-eluting matrix compounds and is thought to take place in the interface. This paper reports strong signal suppression effects which were observed when using a single-stage Orbitrap instrument which was coupled by an electrospray interface to a liquid chromatograph. This type of signal suppression (often the complete loss of certain analyte signal) is observed in addition to signal suppression originating in the electrospray interface. The location of where this phenomenon occurs was shown to be clearly beyond the interface region. It was suspected that not the Orbitrap cell itself, but the C-trap, which is an integral part within the Orbitrap instrument, was the probable location. Such post-interface signal suppression was observed--and could be experimentally induced--when multiply charged ions (e.g. electrospray protonated proteins) were co-eluting with the analytes. High concentrations of proteins, yet not exceeding the maximum ion capacity of the trap, can cause a complete loss of all low m/z masses. This paper describes the practical implication when analyzing heavy matrix samples and discusses strategies to reduce such detrimental effects.
Related Papers
- → A Dual Pressure Linear Ion Trap Orbitrap Instrument with Very High Sequencing Speed(2009)454 cited
- → Imaging MS Methodology for More Chemical Information in Less Data Acquisition Time Utilizing a Hybrid Linear Ion Trap−Orbitrap Mass Spectrometer(2010)84 cited
- → Optimized C-Trap Timing of an Orbitrap 240 Mass Spectrometer for High-Throughput Screening and Native MS by IR-MALDESI(2022)30 cited
- → Applicability of hybrid linear ion trap-high resolution mass spectrometry and quadrupole-linear ion trap-mass spectrometry for mycotoxin analysis in baby food(2011)24 cited
- → Numerical analysis of trajectories in a Cassinian ion trap of second order with trap door ion inlet(2021)4 cited