A simple and robust conductive graphite coating for sheathless electrospray emitters used in capillary electrophoresis/mass spectrometry
Citations Over TimeTop 10% of 2001 papers
Abstract
A graphite-polyimide mixture was used as a conductive coating for sheathless electrospray emitters. The coating procedure described is simple and inexpensive compared to previously described methods. An investigation of the stability of the conductive coating carried out by electrochemical methods revealed good performances during oxidative stress. In addition, no decrease in emitter performance was seen during continuous electrospray in the positive electrospray mode for two weeks. Fast capillary electrophoresis with attomole sensitivity demonstrated the excellent performance of the described sheathless interface when used in conjunction with an orthogonal time-of-flight mass spectrometer. The overall simplicity, stability and low cost of this type of sheathless emitter make the described approach highly suitable for any on-column coupling of low flow rate separation techniques to a mass spectrometer.
Related Papers
- → Twenty years of amino acid determination using capillary electrophoresis: A review(2021)79 cited
- → Recent Progress in the Analysis of Pharmaceuticals by Capillary Electrophoresis(2005)41 cited
- → Electrochemical detection for capillary electrophoresis(1991)54 cited
- Advance in Research of Electrochemiluminescence of Ru(bpy)_3~(2+) Subjected to Capillary Electrophoresis(2010)
- Applications of Chemiluminescence Detection for Capillary Electrophoresis in Drug Analysis(2009)