Attomole-Sensitivity Electrospray Source for Large-Molecule Mass Spectrometry
Analytical Chemistry1995Vol. 67(20), pp. 3802–3805
Citations Over TimeTop 1% of 1995 papers
Abstract
Full mass spectra of high resolving power are obtained from 0.2 nL sample volumes of large (> 10 kDa) nucleotides and proteins using a new electrospray ionization (ESI) system combined with Fourier transform mass spectrometry. The ESI needles are fabricated by laser-heated pulling of fused-silica tubing (5-20 microns i.d.), followed by chemical etching and surface metalization. Total analyte loaded at the instrument of 8.6 fmol and 216 amol produces signal-to-noise ratios of 400:1 and 60:1, respectively, and resolving power of > 10(5) for full mass spectra, while the total amount of material consumed is approximately 150 and 10 amol, respectively.
Related Papers
- → Recent applications of capillary electrophoresis‐electrospray ionisation‐mass spectrometry in drug analysis(2006)76 cited
- → Electrospray mass spectrometry of the self-assembly of a capped polymetallic complex(1993)89 cited
- → Nanospray mass spectrometry in protein and peptide chemistry(2000)10 cited
- Determination of additives, contaminants and byproducts in insulating liquids with desorption electrospray mass spectrometry(2015)
- → Studies on the process of electrospray and applications of electrospray mass spectrometry.(1998)