Tandem Reflectron Time-of-Flight Mass Spectrometer Utilizing Photodissociation
Citations Over TimeTop 22% of 1995 papers
Abstract
A tandem time-of-flight (TOF) mass spectrometer has been designed to obtain complete MS/MS spectra from compounds eluting from a gas chromatograph. This application requires high spectral generation rate, unit mass resolution for both precursor selection and product spectra, and efficient ion utilization. These objectives are achieved by reflectron TOF mass separation in both stages and laser photoinduced dissociation as the ion fragmentation method. Careful timing of the laser pulse relative to ion extraction allows ions of a single m/z value up to m/z 1000 to be photodissociated while ions with adjacent m/z values are essentially unaffected. The convergent foci of the ion packet and laser pulse results in ion fragmentation efficiencies as high as 79%. An ion gate prevents the nonselected precursor ions from convoluting the product spectra. Product spectra can be generated at the maximum laser repetition rate (currently 200 Hz). To achieve unit mass resolution for all product m/z values simultaneously, a novel reflectron was designed for the second TOF stage.
Related Papers
- → Converting a reflectron time-of-flight mass spectrometer into a tandem instrument(1991)21 cited
- → Optimization of a quadrupole ion storage trap as a source for time‐of‐flight mass spectrometry(2012)7 cited
- → A double/triple time-of-flight mass spectrometer for the study of photoprocesses in clusters, or how to produce cluster ions with different temperatures(1991)27 cited
- → High Resolution TOF Secondary Ion Mass Spectrometer(1986)7 cited
- → Time‐of‐Flight Mass Spectrometry(2010)16 cited