Surface-Induced Dissociation on a MALDI-Ion Mobility-Orthogonal Time-of-Flight Mass Spectrometer: Sequencing Peptides from an “In-Solution” Protein Digest
Citations Over TimeTop 10% of 2001 papers
Abstract
Peptide sequencing by surface-induced dissociation (SID) on a MALDI-ion mobility-orthogonal TOF mass spectrometer is demonstrated. SID of approximately 100-fmol amounts of model peptides HLGLAR (m/z 666.8), gramicidin S (m/z 1142.5), and bovine insulin b chain (m/z 3495.5) was accomplished using hydrocarbon-coated gold grids and approximately 20-eV collision energies. The current version of the instrument achieves a mobility resolution of approximately 20 and TOF mass resolution better than 200. Peptide sequences of four peptides from a tryptic digest of cytochrome c (approximately 1 pmol deposited) were obtained. The advantage of IM-SID-o-TOF-MS is that a single experiment can be used to simultaneously measure the molecular weights of the tryptic peptide fragments (e.g., peptide mass mapping) and partial sequence analysis, (e.g., real-time tandem mass spectrometry.)