High-Throughput Nanoliter Sample Introduction Microfluidic Chip-Based Flow Injection Analysis System with Gravity-Driven Flows
Citations Over TimeTop 10% of 2005 papers
Abstract
In this work, a simple, robust, and automated microfluidic chip-based FIA system with gravity-driven flows and liquid-core waveguide (LCW) spectrometric detection was developed. The high-throughput sample introduction system was composed of a capillary sampling probe and an array of horizontally positioned microsample vials with a slot fabricated on the bottom of each vial. FI sample loading and injection were performed by linearly moving the array of vials filled alternately with 50-microL samples and carrier, allowing the probe inlet to enter the solutions in the vials through the slots sequentially and the sample and carrier solution to be introduced into the chip driven by gravity. The performance of the system was demonstrated using the complexation of o-phenanthroline with Fe(II) as a model reaction. A 20-mm-long Teflon AF 2400 capillary (50-microm i.d., 375-microm o.d.) was connected to the chip to function as a LCW detection flow cell with a cell volume of 40 nL and effective path length of 1.7 cm. Linear absorbance response was obtained in the range of 1.0-100 microM Fe(II) (r2=0.9967), and a good reproducibility of 0.6% RSD (n=18) was achieved. The sensitivity was comparable with that obtained using conventional FIA systems, which typically consume 10,000-fold more sample. The highest sampling throughput of 1000 h-1 was obtained by using injection times of 0.08 and 3.4 s for sample and carrier solution, respectively, with a sample consumption of only 0.6 nL for each cycle.
Related Papers
- → Short- and long-term reproducibility of radioisotopic examination of gastric emptying(1990)20 cited
- The reproducibility of reported height and body weight in repeated questionnaire surveys.(1995)
- The reproducibility of scores from three video formats.(1988)
- → Effect of Inorganic Salts on the Determination of Serum Alubmin by Dye-Binding Method with Absorbance Decrease(2007)
- Comparisons of probing depth measurements with Florida probe and conventional periodontal probe(2005)