Variation of the Fourier Transform Mass Spectra Phase Function with Experimental Parameters
Citations Over TimeTop 14% of 2011 papers
Abstract
It has been known for almost 40 years that phase correction of Fourier transform ion cyclotron resonance (FTICR) data can generate an absorption-mode spectrum with much improved peak shape compared to the conventional magnitude-mode. However, research on phasing has been slow due to the complexity of the phase-wrapping problem. Recently, the method for phasing a broadband FTICR spectrum has been solved in the MS community which will surely resurrect this old topic. This paper provides a discussion on the data processing procedure of phase correction and features of the phase function based on both a mathematical treatment and experimental data. Finally, it is shown that the same phase function can be optimized by adding correction factors and can be applied from one experiment to another with different instrument parameters, regardless of the sample measured. Thus, in the vast majority of cases, the phase function needs to be calculated just once, whenever the instrument is calibrated.
Related Papers
- → The phase problem(2003)125 cited
- → Retrieving low- and medium-resolution structural features of macromolecules directly from the diffraction intensities – a real-space approach to the X-ray phase problem(2008)9 cited
- → The phase problem for one-dimensional crystals(2017)13 cited
- → The phase problem for two-dimensional crystals. II. Simulations(2018)1 cited
- → Uniqueness for ab initio phase retrieval in macromolecular X-ray crystallography(2015)