Laser Ionization Orthogonal Time-of-Flight Mass Spectrometry for Simultaneous Determination of Nonmetallic Elements in Solids
Citations Over TimeTop 10% of 2010 papers
Abstract
The simultaneous determination of nonmetallic elements in solid samples is difficult owing to their discrepant physical and chemical properties. We developed a high-irradiance laser ionization orthogonal time-of-flight mass spectrometry (LI-O-TOFMS) system and applied it for the determination of nonmetallic elements in solids. Helium was used as the buffer gas at 250 Pa in the source chamber; the laser irradiance was about 7 x 10(10) W/cm(2). A series of artificial standards containing B, C, N, O, F, Si, P, S, Cl, As, Br, Se, I, and Te were used. Explicit spectra were obtained with only a little interference from gas species and doubly charged matrix ions. Standardless semiquantitative analysis could be accomplished with a novel sampling methodology to obtain near-uniform sensitivity coefficients for different elements. Limits of detection (LOD) at microgram per gram level and a dynamic range of 6 orders of magnitude were achieved for most nonmetallic elements.
Related Papers
- → Collisional activation mass spectra(1979)40 cited
- → Measurement of nanoparticle mass distributions by laser desorption/ionization time-of-flight mass spectrometry(2007)4 cited
- → Time‐of‐Flight Mass Spectrometry(2010)16 cited
- → Chemical ionization Fourier transform mass spectrometry of chemical warfare agent simulants using laser-produced metal ions(1990)5 cited
- → Study of C22X- Organic Salts by Simultaneous Use of Fast Atom Bombardment Mass Spectrometry and Tandem Mass Spectrometry. Chemical Reactivity in the Gaseous Phase(1997)1 cited