A Quadrupole Ion Trap with Cylindrical Geometry Operated in the Mass-Selective Instability Mode
Citations Over TimeTop 13% of 1998 papers
Abstract
A cylindrical geometry ion trap is used to record mass spectra in the mass-selective instability mode. The geometry of the cylindrical ion trap has been optimized to maximize the quadrupole field component relative to the higher-order field content through field calculations using the Poisson/Superfish code and through experimental variation of the electrode structure. The results correspond well with predictions of the calculations. The trap has been used to record mass spectra with better than unit mass resolution, high sensitivity, and a mass/charge range of ∼600 Th. Multistage (MS(3)) experiments have been performed, and the Mathieu stability region has been experimentally mapped. The performance of this device compares satisfactorily with that of the hyperbolic ion trap.