DNA Sequencing on Microfabricated Electrophoretic Devices
Citations Over TimeTop 1% of 1998 papers
Abstract
We present a model that quantitatively describes the performance of microfabricated electrophoretic devices filled with linear polyacrylamide as replaceable sieving material for single-stranded DNA analyses. The dependence of resolution on various separation parameters such as selectivity, diffusion, injector size, device length, and channel folding was investigated. A previously predicted dependence of longitudinal diffusion coefficient on electric field strength has been verified. We have used this model to develop and optimize microfabricated electrophoretic devices for DNA analyses. For single-color DNA sequencing mixtures, we routinely achieve separations of 400 bases in under 14 min at 200 V/cm, and separation of 350 bases in only 7 min at 400 V/cm, with a minimum resolution of R = 0.5. Our results also indicate reduced fragment biasing and efficient sample stacking for DNA sample loading on microfabricated devices.
Related Papers
- → SDS-assisted protein transport through solid-state nanopores(2017)82 cited
- → Slowing single-stranded DNA translocation through a solid-state nanopore by decreasing the nanopore diameter(2014)77 cited
- → Selectively detecting attomolar concentrations of proteins using gold lined nanopores in a nanopore blockade sensor(2020)36 cited
- → Effect of nanopore size on poly(dT)30 translocation through silicon nitride membrane(2013)22 cited
- → Survey of possible layer stacking structures*(1967)25 cited