Graphene-based carbon-layered electrode array technology for neural imaging and optogenetic applications
Citations Over TimeTop 1% of 2014 papers
Abstract
Neural micro-electrode arrays that are transparent over a broad wavelength spectrum from ultraviolet to infrared could allow for simultaneous electrophysiology and optical imaging, as well as optogenetic modulation of the underlying brain tissue. The long-term biocompatibility and reliability of neural micro-electrodes also require their mechanical flexibility and compliance with soft tissues. Here we present a graphene-based, carbon-layered electrode array (CLEAR) device, which can be implanted on the brain surface in rodents for high-resolution neurophysiological recording. We characterize optical transparency of the device at >90% transmission over the ultraviolet to infrared spectrum and demonstrate its utility through optical interface experiments that use this broad spectrum transparency. These include optogenetic activation of focal cortical areas directly beneath electrodes, in vivo imaging of the cortical vasculature via fluorescence microscopy and 3D optical coherence tomography. This study demonstrates an array of interfacing abilities of the CLEAR device and its utility for neural applications.
Related Papers
- → Functional ultrasound imaging of the spreading activity following optogenetic stimulation of the rat visual cortex(2021)16 cited
- → Application of Optogenetics in Gene Therapy(2018)4 cited
- Optogenetic cardiac pacing and its applications(2021)
- → Optogenetic Cardiac Pacing and Its Applications(2021)
- → The effect of optogenetic activation of astrocytes on the hippocampal neurons activity(2021)