Wide-field tissue polarimetry allows efficient localized mass spectrometry imaging of biological tissues
Citations Over TimeTop 10% of 2015 papers
Abstract
While mass spectrometers can detect chemical signatures within milliseconds of data acquisition time, the non-targeted nature of mass spectrometry imaging (MSI) necessitates probing the entire surface of the sample to reveal molecular composition even if the information is only sought from a sample subsection. This leads to long analysis times. Here, we used polarimetry to identify, within a biological tissue, areas of polarimetric heterogeneity indicative of cancer. We were then able to target our MS analysis using polarimetry results to either the cancer region itself or to the cancer margin. A tandem of polarimetry and Desorption Electrospray Ionization Mass Spectrometry Imaging (DESI-MSI) enables fast (10 fold compared to non-targeted imaging), and accurate pathology assessment (cancer typification in less than 2 minutes compared to 30 minutes for histopathology) of ex vivo tissue slices, without additional sample preparation. This workflow reduces the overall analysis time of MSI as a research tool.
Related Papers
- → Imaging of Metabolites by MALDI Mass Spectrometry(2009)49 cited
- → The application of mass spectrometry imaging in traditional Chinese medicine: a review(2022)27 cited
- → Mass spectrometry imaging: Facts and perspectives from a non-mass spectrometrist point of view(2012)16 cited
- → Metabolite imaging by mass spectrometry: A new discovery tool(2020)5 cited
- → Matrix-Assisted Laser Desorption Ionization Imaging Mass Spectrometry of Drug Distribution in Mouse Brain Tissue by High-Resolution Time-of-Flight Mass Spectrometry(2018)3 cited