Expanding the momentum field of view in angle-resolved photoemission systems with hemispherical analyzers
Review of Scientific Instruments2021Vol. 92(12), pp. 123907–123907
Citations Over Time
N. Gauthier, Jonathan A. Sobota, Heike Pfau, Alexandre Gauthier, Hadas Soifer, Maja D. Bachmann, I. R. Fisher, Zhi‐Xun Shen, P. S. Kirchmann
Abstract
In photoelectron spectroscopy, the measured electron momentum range is intrinsically related to the excitation photon energy. Low photon energies 3. By providing a larger field of view in momentum space, our approach using a bias potential considerably expands the flexibility of laser-based photoemission setups.
Related Papers
- → Increasing a spectrometer's solid angle without loss of resolution(1977)10 cited
- → Construction of an electron analyzer with a wide acceptance solid angle(2005)5 cited
- MAGNETIC ANALYZER FOR STUDIES OF NUCLEAR REACTIONS(1958)
- → A Large-Solid-Angle Electron Energy Analyzer.(2001)