Physical Working Principles of Medical Radar
Citations Over TimeTop 10% of 2012 papers
Abstract
There has been research interest in using radar for contactless measurements of the human heartbeat for several years. While many systems have been demonstrated, not much attention have been given to the actual physical causes of why this work. The consensus seems to be that the radar senses small body movements correlated with heartbeats, but whether only the movements of the body surface or reflections from internal organs are also monitored have not been answered definitely. There has recently been proposed another theory that blood perfusion in the skin could be the main reason radars are able to detect heartbeats. In this paper, an experimental approach is given to determine the physical causes. The measurement results show that it is the body surface reflections that dominate radar measurements of human heartbeats.
Related Papers
- → Local Pulse Wave Detection Using Continuous Wave Radar Systems(2017)57 cited
- → Heartbeat and Respiration Detection Using a Low Complexity CW Radar System(2021)11 cited
- → Radar for healthcare technology(2022)1 cited
- → A distributed multi MIMO radar imaging system(2022)