Comparison of Edwards Sapien XT versus Lotus Valve Devices in Terms of Electrophysiological Study Parameters in Patients Undergoing TAVI
Pacing and Clinical Electrophysiology2016Vol. 39(10), pp. 1132–1140
Citations Over TimeTop 23% of 2016 papers
Abdurrahman Eksik, Aydın Yildirim, Mehmet Gül, Serkan Aslan, Aydın Rodi Tosu, Özgür Sürgit, Hüseyin Altuğ Çakmak, Muhammet Hulusi Satılmışoğlu, Emre Akkaya, İhsan Bakır
Abstract
In this first study comparing the findings of EPS and electrocardiography, the impact of the Lotus valve on AV conduction systems was greater than that of the Sapien XT. However, the need for PPM was higher in the Lotus valve than in the Sapien XT. PPM requirement is related to valve design; it may decrease with reduced frame height and metal burden in novel valve systems.