A Comprehensive Evaluation of the Precision (Repeatability and Reproducibility) of the Oculus Pentacam HR
Investigative Ophthalmology & Visual Science2011Vol. 52(10), pp. 7731–7731
Citations Over TimeTop 1% of 2011 papers
Abstract
Measurements taken with the Pentacam HR are repeatable and reproducible, especially those obtained with the cornea fine scan. Although the Pentacam HR is clearly a very useful clinical and research tool, the measurement of corneal axes, pupil center pachymetry, front meridional and axial maps, refractive power maps, and equivalent K readings should be interpreted with caution.