Biomechanical Responses of Healthy and Keratoconic Corneas Measured Using a Noncontact Scheimpflug-Based Tonometer
Investigative Ophthalmology & Visual Science2014Vol. 55(6), pp. 3651–3651
Citations Over TimeTop 1% of 2014 papers
Abstract
Key parameters assessed by the CST are repeatable. Keratoconus is associated with greater DA than in healthy eyes, even when controlled for CCT and IOP. Deformation amplitude may be a useful adjunct in keratoconus assessment and monitoring, but cannot solely discriminate between healthy and keratoconic corneas.