Evaluation of National Institute of Standards and Technology (NIST) Thermal Conductivity Standard Reference Materials Using a 20-cm Heat Flow Meter Apparatus
Journal of Thermal Insulation1989Vol. 13(1), pp. 62–70
Abstract
The equations developed for NIST (formerly National Bureau of Standards) thermal conductivity Standard Reference Materials (SRM's) were derived from thermal conductivity measurements made on guarded hot plate apparatuses at NIST. These instruments are briefly described in the certificates and more fully described in NIST special publications [1,2] Critical to the conclusions ofthis report is the size of the heat flow meter apparatus used for making measurements on the NIST thermal conductivity standards. This work describes additional variability in the NIST thermal conductivity standards when they are measured on a system which is smaller than the guarded hot plate apparatuses used at NIST
Related Papers
- → NIST–IMGC comparison of gas flows below one litre per minute(2003)18 cited
- → International comparison of a NIST primary standard with an NRLM transfer standard for small mass flow rates of nitrogen gas(1998)13 cited
- Updated NIST standards for dose calibrator measurements of imaging and therapeutic nuclides(2014)
- → An improved, optically-pumped, primary frequency standard(2002)2 cited
- NIST-7, The U.S. Primary Frequency Standard: New Evaluation Techniques(1996)