Capture and Analysis of Low Molecular Weight Ligands by Surface Plasmon Resonance Combined with Mass Spectrometry
Citations Over TimeTop 18% of 2001 papers
Abstract
The combination of biomolecular interaction analysis (BIA)by surface plasmon resonance (SPR)and nano-electrospray ionization ion trap mass spectrometry (nanoESI-Ion Trap MS)as well as matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-ToF MS)is demonstrated for the binding of low molecular weight inhibitors (∼ 600 Da) to HIV-1 protease. Inhibitors were captured on sensor chips of a manual or an automated SPR biosensor, to which HIV-1 protease was immobilized. Compounds and buffer components that bound unspecifically to the sensor surface were removed and the inhibitors were eluted in a minimal volume (3 μL), between air bubbles, in order to prevent dispersion of analyte into buffer eluent. Molecular weights were subsequently determined by mass spectrometry, structural information was obtained by MALDI-ToF post-source decay as well as by electrospray ionization tandem mass spectrometry (MS/MS)analysis. Furthermore, competition experiments, using a mixture of different ligands, demonstrated that the peak intensities in the MALDI-ToF spectrum could be used for relative quantification of the amount of the different ligands bound to the immobilized target. Methodology for automated capture and elution of analytes was developed and evaluated.
Related Papers
- → Analytical Characterization of the Electrospray Ion Source in the Nanoflow Regime(2008)90 cited
- → Synchronized dual-polarity electrospray ionization mass spectrometry(2009)8 cited
- → Suppression of Protein Structural Perturbations in Native Electrospray Ionization during the Final Evaporation Stages Revealed by Molecular Dynamics Simulations(2021)11 cited
- → Electro-kinetic assisted electrospray ionization for enhanced complex sample analysis(2016)10 cited
- → Pneumatically assisted electrospray-ion mobility spectrometry for quantitative analysis of intact proteins(2012)7 cited