Comparison of exhalation time methods (6 sec vs. 10 sec) of a hand‐held exhaled nitric oxide analyzer
Citations Over TimeTop 22% of 2010 papers
Abstract
Standard exhalation time for measuring fractional exhaled nitric oxide (FeNO) is 10 sec, but this is not easy for younger children. We aimed to investigate the agreement between FeNO values during 10-sec (FeNO-10) and 6-sec (FeNO-6) exhalation and the feasibility of measuring FeNO-6, using a hand-held analyzer, NIOX-MINO®. FeNO values measured during 10- and 6-sec (random order) were compared. Success rates of the two different time modes were also evaluated. In 119 asthmatic children (median age 8 years [range 4-15]) who had been already accustomed to NIOX-MINO®, median FeNO-10 (29 ppb [IQR 15.2-42.0]) and FeNO-6 (27 ppb [IQR 16.0-43.5]) did not differ significantly (P = 0.90), and there was a good correlation between both values (r = 0.984, P < 0.001). Mean difference (FeNO-10-FeNO-6) was -0.151 ppb (95% CI: -0.95 to 0.65, limits of agreement: -8.8 to 8.5). In 46 asthmatic children (median age 7 years [range 4-15]) who had never used any FeNO analyzers, all the children aged 8 years and more (n = 21) succeeded in measuring FeNO on both time modes, whereas for children aged younger than 8 years (n = 25) success rates of the 10- and 6-sec mode were 60.0% and 92.0%, respectively. In conclusion, we showed good agreement between FeNO-10 and FeNO-6, and the 6-sec mode of NIOX-MINO® is more feasible than 10-sec mode for measuring FeNO in younger children.
Related Papers
- → Comparison of exhalation time methods (6 sec vs. 10 sec) of a hand‐held exhaled nitric oxide analyzer(2010)14 cited
- → Successful Fractional Exhaled Nitric Oxide Measurements with NIOX VERO using 6- and 10-s Exhalation Modes in Children with Asthma Aged 4–10 Years: A Randomized Study(2017)6 cited
- The measurement of exhaled nitric oxide in routine practice.(2009)
- → Agreement Between the 6 Seconds and 10 Seconds Exhalation Modes When Measuring Fractional Exhaled Nitric Oxide in Children Aged 4-6 Years With the NIOX VERO Device(2017)
- → P214 CLINICAL PRECISION, ACCURACY, NUMBER AND DURATION OF EXHALATIONS FOR A NOVEL EXHALED NITRIC OXIDE DEVICE(2019)