Use of a lung model to assess mechanical in‐exsufflator therapy in infants with tracheostomy
Pediatric Pulmonology2010Vol. 46(3), pp. 211–217
Citations Over Time
Amanda M. Striegl, Gregory J. Redding, Robert M DiBlasi, Dave Crotwell, John W Salyer, Edward R. Carter
Abstract
Using the MIE via tracheostomy tube in an infant lung model, we found that an insufflation time of >1 sec is required for equilibration of insufflation pressure and alveolar pressure. Longer exsufflation time does not significantly alter MEF. Higher insufflation and exsufflation pressures both increased MEF, but greater exsufflation pressure had more substantial impact.