Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer
Citations Over TimeTop 1% of 2006 papers
Abstract
Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/s) with nominal mass resolving power of 60,000 and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds 100,000 (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second.
Related Papers
- → Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer(2006)751 cited
- → A Dual Pressure Linear Ion Trap Orbitrap Instrument with Very High Sequencing Speed(2009)454 cited
- → Applicability of hybrid linear ion trap-high resolution mass spectrometry and quadrupole-linear ion trap-mass spectrometry for mycotoxin analysis in baby food(2011)24 cited
- → Numerical analysis of trajectories in a Cassinian ion trap of second order with trap door ion inlet(2021)4 cited
- → Coupling ESI and MALDI Sources to the Quadrupole Mass Filter, Quadrupole Ion Trap, Linear Quadrupole Ion Trap, and Orbitrap Mass Analyzers(2010)1 cited